Emerson Model 1700 User Manual

Page of 350
Example: Default display variable configuration
Display variable
Process variable assignment
Display Variable 1
Mass flow
Display Variable 2
Mass total
Display Variable 3
Volume flow
Display Variable 4
Volume total
Display Variable 5
Density
Display Variable 6
Temperature
Display Variable 7
External pressure
Display Variable 8
Mass flow
Display Variable 9
None
Display Variable 10
None
Display Variable 11
None
Display Variable 12
None
Display Variable 13
None
Display Variable 14
None
Display Variable 15
None
Configure Display Variable 1 to track the primary mA output
Display (standard)
OFF-LINE MAINT
 > OFF-LINE CONFG > DISPLY > VAR 1
Chinese-language
display
Offline Maintain
 > Configuration > Display > Configure 1st Var
ProLink II
ProLink
 > Configuration > Display > Var1
ProLink III
Device Tools
 > Configuration > Transmitter Display > Display Security
Field Communicator Configure > Manual Setup > Display > Display Variables
Overview
You can configure Display Variable 1 to track mA Output Process Variable for the primary mA
output. When tracking is enabled, you can control Display Variable 1 from the display menu.
Tip
This feature is the only way to configure a display variable from the display menus (standard display
option only), and it applies only to Display Variable 1.
Procedure
Configure Display Variable 1 to track the primary mA output.
Configure device options and preferences
60
 Micro Motion
®
 Model 1700 Transmitters with Analog Outputs