Teledyne 460L User Manual
TELEDYNE INSTRUMENTS
Getting Started
460L Instruction Manual
16
05228 Rev B
PRINTED DOCUMENTS ARE UNCONTROLLED
DCN 5164
Below each relay is a 3-pin connector that allows the relay to be connected for either
normally open or normally closed operation.
Table 3-4 describes how to connect the alarm relays.
NOTE
The relay contacts are rated to 3A at 240VAC. Do not exceed these ratings when
connecting equipment to the instrument.
RELAY 1
N.
O
.
CO
M
N.
C
.
D6
RELAY 2
N.
O.
CO
M
N.
C
.
D7
RELAY 3
N.
O.
CO
M
N.
C
.
D8
STATUS LEDS
Figure 3-10 Alarm Output Relays
Table 3-4 Relay Output Operation
RELAY PIN
STATE
2
RELAY
FUNCTION
N.
O.
C
O
M
M
N.
C.
C.
STATUS
LED
1
COMMENTS
SENSOR OK
ON
D6 ON
460L is operating normally
1
SENSOR OK
OFF
D6 OFF
Problem with the O
3
sensor module.
See Section 11.2.
HI Alarm
ON
D7 ON
O
3
concentration > HI alarm limit
2
HI Alarm
OFF
D7 OFF
O
3
concentration < HI alarm limit
HI-HI Alarm
ON
D8 ON
O
3
concentration > HI-HI alarm limit
3
HI-HI Alarm
OFF
D8 OFF
O
3
concentration < HI-HI alarm limit