Getting Started 
460L Instruction Manual 
16 
05228 Rev B 
 
PRINTED DOCUMENTS ARE UNCONTROLLED 
DCN 5164 
Below each relay is a 3-pin connector that allows the relay to be connected for either 
normally open or normally closed operation.   
Table 3-4 describes how to connect the alarm relays.   
NOTE  
The relay contacts are rated to 3A at 240VAC.  Do not exceed these ratings when 
connecting equipment to the instrument. 
 
RELAY 1
N.
O
.
CO
M
N.
C
.
D6
RELAY 2
N.
O.
CO
M
N.
C
.
D7
RELAY 3
N.
O.
CO
M
N.
C
.
D8
STATUS LEDS
 
Figure 3-10  Alarm Output Relays 
 
Table 3-4 Relay Output Operation 
RELAY PIN 
STATE
2 
RELAY 
FUNCTION
 
N. 
O.
 

N. 
C. 
STATUS 
LED
1
 
COMMENTS 
SENSOR OK 
ON
 
 
 
 
D6 ON 
460L is operating normally  
SENSOR OK 
OFF
 
 
 
 
D6 OFF 
Problem with the O
3
 sensor module.   
See Section 11.2. 
HI Alarm 
ON
 
 
 
 
D7 ON 
O
3
 concentration > HI alarm limit 
HI Alarm 
OFF
 
 
 
 
D7 OFF 
O
3
 concentration < HI alarm limit 
HI-HI Alarm 
ON
 
 
 
 
D8 ON 
O
3
 concentration > HI-HI alarm limit 
HI-HI Alarm 
OFF
 
 
 
 
D8 OFF 
O
3
 concentration < HI-HI alarm limit