MegaZoom Concepts and Oscilloscope Operation
To setup the Analog channels
Vertical sensitivity  Turn the large knob above the channel key to set the 
sensitivity (volts/division) for the channel. The volts/division knob changes the 
analog channel sensitivity from 1 mV to 5 V for the 54620-series and from 2 mV 
to 5 V for the 54640-series in a 1-2-5 step sequence with a 1:1 probe attached. 
The analog channel Volts/Div value is displayed in the status line.
Vertical expansion  The default mode for expanding the signal when you turn 
the volts/division knob is vertical expansion about the ground level.  To set the 
expansion mode to expand about the center of the screen instead, press 
Expand 
in the 
Utility Options menu and select Center.
Ground level  The ground level of the signal for each displayed analog channel 
is identified by the position of the 
 icon at the far-left side of the display.
Channel position  Turn the small channel position knob (
) to move the 
channel waveform up or down on the display. The voltage value momentarily 
displayed in the upper right portion of the display represents the voltage 
difference between the center graticule and the ground level (
) icon.  It also 
represents the voltage at the center graticule.
2 Press the 
Coupling
 softkey to select the input channel coupling.
Coupling changes the channel’s input coupling to either 
AC (alternating 
current), 
DC (direct current), or Ground (hardware ground on 54620-series 
only). AC coupling places a 3.5 Hz high-pass filter in series with the input 
waveform that removes any DC offset voltage from the waveform. When 
AC is 
selected, “AC” is illuminated on the front panel next to the channel position 
knob (
).
• DC coupling is useful for viewing waveforms as low as 0 Hz that do not 
have large DC offsets.
• AC coupling is useful for viewing waveforms with large DC offsets, or with 
slowing moving DC offsets.
• Ground coupling (54620-series only) is useful for removing clutter on 
screen, estimating the baseline noise contribution by the oscilloscope, and 
measuring the difference between oscilloscope ground and 
device-under-test ground.