To use the Roll horizontal mode
To use the Roll horizontal mode 
• Press the 
Main/Delayed
 key, then press the 
Roll
 softkey.
Roll mode causes the waveform to move slowly across the screen from right to 
left. It only operates on time base settings of 500 ms/div and slower. If the 
current time base setting is faster than the 500 ms/div limit, it will be set to 
500 ms/div when Roll mode is entered.
In Normal horizontal mode, signal events occurring before the trigger are 
plotted to the left of the trigger point (▼) and signal events after the trigger 
plotted to the right of the trigger point.
In Roll mode there is no trigger.  The fixed reference point on the screen is the 
right edge of the screen and refers to the current moment in time.  Events that 
have occurred are scrolled to the left of the reference point.  Since there is no 
trigger, no pretrigger information is available.
If you would like to clear the display and restart an acquisition in Roll mode, 
press the 
Single key.
Use Roll mode on low-frequency waveforms to give a display much like a strip 
chart recorder.  It allows the waveform to roll across the display.