This chapter describes the basic operation of the Ultrafine Particle 
Monitor - Model 651 and describes how to use the controls, 
indicators, and connectors found on the front and back panels.  
O p e r a t i n g   P r e c a u t i o n s  
Read the following before applying power to the particle monitor:  
  Review the operating specifications for the Model 651 described 
in 
Appendix A
  Do not operate the Model 651 at temperatures outside the 
range of 10C to 35C. If the particle monitor is operated outside 
this range, the displayed concentration may be inaccurate. 
 
 
W A R N I N G
 
  The Model 651 should not be used with hazardous gases such as 
hydrogen or oxygen. Using the particle monitor with hazardous gases 
may cause injury to personnel and damage to equipment. 
R e c o m m e n d e d   O p e r a t i o n   P r o c e d u r e s  
Outdoor Operation Procedures 
When sampling outdoor aerosol, follow these recommendations: 
  Place the Model 651 in a conditioned enclosure or shelter to 
ensure that it is operating within temperature and humidity 
specifications. 
  If the Model 651 is placed in an environment with temperatures 
lower than the ambient temperature, consider heating the 
sample line to reduce condensation.  
  If you are not using a sampling system, use a cyclone with a cut 
size no greater than 3 µm on the particle counter inlet.  
  Ensure that the pressure differential at the inlet is not greater 
than 2.5 kPa (10 inches of H
2
O). If you are using a cyclone, do 
not exceed the inlet pressure drop of 2.5 kPa.