Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
3
Methods
 
To setup a method to use the multimode source
To setup a method to use the multimode source
To have your method use a multimode source, follow these steps:
Open the MSD Spray Chamber dialog box by clicking Instrument > MSD 
Spray Chamber in the Method and Run Control view.
Set Method Spray Chamber to MM-ES+APCI.
Verify that Installed Spray Chamber is set to MM-ES+APCI.
Make any other changes that are necessary for your method.
Click the OK button.
Open The Set up MSD Signals dialog box by clicking Instrument > More > 
Set up MSD Signals in the Method and Run Control view.
Choose the desired ionization mode from the Ionization list. This list is 
only visible if the Method Spray Chamber was set to MM-ES+APCI. You 
may set the ionization mode to one of the following:
MM-ES
MM-APCI
MM-ES+APCI
Make any other changes that are necessary for your method.
Click the OK button.
WA R N I N G
The 6100 Series Single Quad LC/MS Liquid Chromatograph diverter valve is an 
integral part of the G1978B safety system. The LC mobile phase flow must always be 
connected to the diverter valve inlet filter. Never bypass the diverter valve and 
connect directly to the nebulizer. If the diverter valve is used in a manner not 
specified by Agilent Technologies, the protections provided by the diverter valve may 
be impaired and the system may catch fire.