A&D AD-4212A/B User Manual

Page of 102
 
32
 
Class Item 
and 
Parameter 
Description 
 0 
 600 
bps 
 1 
 1200 
bps 
 2 
 2400 
bps 
 3 
 4800 
bps 
 4 
 9600 
bps 
bp5
 
Baud rate
 
 5 
 19200 bps 
 
 0 
7 bits, even 
 1 
7 bits, odd 
btpr
 
Data bit, parity bit
 
 2 
8 bits, none 
 
 0 
CR LF 
Crlf
 
Terminator
 
 1 
CR 
CR: ASCII code 0Dh   
LF: ASCII code 0Ah 
 0 
A&D standard format 
 1 
DP format 
 2 
KF format 
 3 
MT format 
 4 
NU format 
type
 
Data format
 
 5 
CSV format 
Refer to "10-6 Description of 
Item "Data Format". 
 0 
No limit 
t-Up
 
Timeout
 
 1 
1 second 
Selects the wait time to receive a 
command. 
 0 
No output 
erCd
 
AK, Error code
 
 1 
Output 
AK: ASCII code 06h 
 0 
Not used 
 
 
 
 
 
 
 
5if 
Serial 
interface 
 
Not displayed 
when BCD 
output (OP-01) 
is installed. 
 
 
Ct5
 
CTS, RTS control
 
 1 
Used 
Controls CTS and RTS. 
 0 
OFF at 1 ON 
at 
datp
 
Data 
 1 
ON at 1 OFF 
at 
 0 
OFF at negative 
ON at positive or zero 
polp
 
Polarity 
  1 
ON at negative 
OFF at positive or zero 
 0 
OFF at stabilization indicator ON    ON at stabilization indicator OFF 
5tbp
 
Stability 
  1 
ON at stabilization indicator ON        OFF at stabilization indicator OFF 
 0 
OFF at e, -e display 
0erp
 
OVER 
  1 
ON at e, -e display 
 0 
Data refresh completed when ON
→OFF 
5trp
 
STROBE 
  1 
Data refresh completed when OFF
→ON 
 0 
OFF when weighing 
ON when not weighing *4 
 
 
 
bcd 
 
 
Displayed only 
when BCD 
output (OP-01) 
is installed. 
 
 
 
5tap
 
Status 
  1 
ON when weighing 
OFF when not weighing *4 
Unit 
Unit
 
Refer to “5. WEIGHING UNITS”. 
id 
ID number setting
 
Refer to “11. ID NUMBER AND GLP REPORT”. 
erfnc 
Extended functions
 
Available only for the AD-4212A series balance when “To use 
the extended functions” is selected for the function switch and is 
used for special applications.   
For details, refer to “20. EXTENDED FUNCTIONS”. 
           Factory setting 
 
*4    “Not weighing” is a condition that a weight value is not displayed such as re-zero operation or 
display-off state. 
Caution 
The balance may not transmit the data completely at the specified refresh rate, depending on 
the baud rate or data added to the weighing data such as time, date or ID number.