Philips M1191AL User Manual

Page of 247
8
Cleaning and Low-Level Disinfection
The reusable sensors should be cleaned and disinfected, but never 
sterilized. Follow the procedure outlined below. 
Warnings
Use only validated cleaning agents and disinfectants listed below; do 
not use any others. Failure to do so may damage the sensor or its con-
necting wires and shorten product lifetime or cause safety hazards.
Select disinfectants carefully as some have very similar names but 
completely different compositions.
Do not immerse the sensor connector in any of the cleaning solutions, 
disinfectant, or other liquid (only the sensor and cable housing may be 
immersed, not the connector).
Do not soak sensors in disinfectants longer than specified by the dis-
infectant manufacturer.
Do not sterilize the sensors.
Validated Cleaning Agents
Validated Disinfectants
- Mild Detergent
- Salt Solution (1%)
Metricide
®
 28
Cidex
®
 Formula 7
Kohrsolin
®
 (2%)
Metricide
®
 Plus 30
Cidex
®
 OPA
Mucapur
®
-CD (1%)
Terralin
®
 Liquid
Cidex
®
 Plus
Isopropanol (70%) or
Isopropanol Wipe (70%)
Incidin
®
 Liquid
Omnicide
®
 28
Step
Cleaning and Low-Level Disinfection
1
Clean sensor following instructions supplied with cleaning agent. 
2
Disinfect sensor following instructions supplied with disinfectant.
3
Rinse sensor in water, wiping it dry with a clean cloth, leaving to dry com-
pletely. If you notice any deterioration or damage to sensor or cable, dis-
pose the sensor immediately.