Honeywell SC1000 User Manual

Page of 174
 page 72      008-0608-00
applied this load to the transducer, press [ENTER]. 
Upon completion, the display will indicate 
DONE
 and the instrument will return to the 
RUN mode.
If the 
CALIBRATION TYPE
 is 5-Point Known Load Calibration...
The display will read 
DOING 5POINT CAL
, and prompt you to 
APPLY 00000 UNIT
 (where 
00000 and UNIT are previously entered as
 KNOWN POINT 1/5
 and 
DISPLAY. UNITS
).  
When you have applied this load to the transducer, press [ENTER]. 
The display will read 
WORKING
, then 
APPLY 00250 UNIT
 (where 00250 and UNIT are pre-
viously entered as
 KNOWN POINT 2/5
 and 
DISPLAY. UNITS
). When you have applied this 
load to the transducer, press [ENTER].
The display will read 
WORKING
, then 
APPLY 00500 UNIT
 (where 00500 and UNIT are pre-
viously entered as
 KNOWN POINT 3/5
 and 
DISPLAY. UNITS
). When you have applied this 
load to the transducer, press [ENTER].
The display will read 
WORKING
, then 
APPLY 00750 UNIT
 (where 00750 and UNIT are pre-
viously entered as
 KNOWN POINT 4/5
 and 
DISPLAY. UNITS
). When you have applied this 
load to the transducer, press [ENTER].
The display will read 
WORKING
, then 
APPLY 01000 UNIT
 (where 01000 and UNIT are pre-
viously entered as
 KNOWN POINT 5/5
 and 
DISPLAY. UNITS
). When you have applied this 
load to the transducer, press [ENTER].
The display will now read 
RETURN PNTS.=NO
. If you press [ENTER], the display will 
indicate 
DONE
 and the instrument will return to the RUN mode.  If you use the [UP] 
button and [ENTER] to select 
RETURN POINTS=YES
, you may now apply the same 
loads again but in reverse order.  This will allow the instrument to measure and 
partially eliminate any hysteresis that may be present in the transducer.  After the 
loads have been applied again, the display will indicate 
DONE
 and the instrument 
will return to the RUN mode.