Emerson 228 User Manual

Page of 16
Model 228
Submersion/Insertion Toroidal Sensor
Instruction Sheet
PN 51A-228/rev.P
December 2010
SPECIFICATIONS - SENSORS
MODEL
Wetted Materials
Temperature  
Pressure                    CRN pressure*
228-02               PEEK (glass-filled)
248°F (120°C)
295 psig (2135 kPa abs)
220 psig (1618 kPa abs)
228-03               PEEK (glass-filled)
392°F (200°C)
295 psig (2135 kPa abs)
220 psig (1618 kPa abs)
228-04               Tefzel (glass-filled)
248°F (120°C)        200 psig (1480 kPa abs)
150 psig (1135 kPa abs)
228-05                Tefzel (unfilled)
248°F (120°C)        200 psig (1480 kPa abs)
150 psig (1135 kPa abs)
SPECIFICATIONS - ADAPTERS
SPECIFICATIONS - RETRACTION ASSEMBLIES
Operating Conditions
Retraction/Insertion Conditions
PN
Wetted Materials
Temperature
Pressure
Temperature
Pressure
23311-00 
392°F 
295 psig
392°F 
295 psig
(mechanical)
(200°C)
(2135 kPa abs)
(200°C)
(2135 kPa abs)
23311-01 
316 SS, Teflon, EPDM
392°F
295 psig
266°F
35 psig
(manual)
(200°C)
(2135 kPa abs)
(130°C)
(343 kPa abs)
PN
Wetted Materials            Temperature
Pressure                    CRN pressure*
23242-02
PEEK, Viton, 316 SS           392°F (200°C)
295 psig (2135 kPa abs)       220 psig (1618 kPa abs)
23242-03            PEEK, Viton, 316 SS           392°F (200°C)
295 psig (2135 kPa abs)       220 psig (1618 kPa abs)
2001990              CVPC, PEEK, Viton            100°F (38°C)
100 psig (791 kPa abs)                       NA
185°F (85°C)
45 psig (412 kPa abs)
For additional information, please visit our website 
at www.emersonprocess.com/raihome/liquid/.
Maximum pressure for applications requiring CRN registration
Maximum pressure for applications requiring CRN registration
316 SS, Teflon, EPDM
316 SS, Teflon, EPDM
NA
CVPC, PEEK, Viton   
2001990    
WARNING
Before removing the sensor, be absolutely certain
that the process pressure is reduced to 0 psig and
the process temperature is lowered to a safe level!
CAUTION
The wetted sensor materials may not be compatible
with process com position and operating conditions.
Application compat ibility is entirely the responsi-
bility of the user.
CAUTION
SENSOR/PROCESS
APPLICATION COMPATIBILITY