Honeywell SC2000 User Manual

Page of 174
 
SC Series Instruction Manual     page 107
Hig
h-L
ev
el
 In
pu
t C
han
nel
12
12.4  Calibration Procedure
If you are not familiar with operating the instrument in the SETUP menu mode,
see “SETUP Menu mode” on page 23.  A listing of all menu items is given in
“Setup Menu Reference” on page 157.
Step 1: Wire the transducer to the channel’s connector.
Step 2: Set the Excitation and Signal jumpers appropriate for the transducer,
Step 3: Enter the 
CALIBRATION TYPE
.  
There are two methods that can be used to calibrate the transducer to the
Input Channel.  Each has advantages and disadvantages as described in
“CALIBRATION TYPE Menu Item” on page 113. It is important to know your
application in order to select the appropriate calibration type.
Step 4: Enter the
 CALIBRATION DATA.
Otherwise, consult the Certificate of Calibration for the transducer when
entering information in the 
CALIBRATION DATA
 sub-menu.
Step 5: Perform the calibration.
Using the 
CALIBRATE
 menu item starts the calibration process.  You will be
prompted to apply loads to the transducer as required.