Emerson 381 User Manual
17
Trouble
Probable Cause
Remedy
Meter reads off scale.
Reference electrolyte depleted.
Replace reference electrolyte as
instructed in Section 3.2.2.
instructed in Section 3.2.2.
Defective preamplifier.
Check preamplifier as instructed in
Section 3.4 and replace
preamplifier if defective.
Section 3.4 and replace
preamplifier if defective.
T.C. element shorted.
Check T.C. element as instructed in
Section 3.5, and replace body if
defective.
Section 3.5, and replace body if
defective.
Electrode not in solution or sample
Make sure sensor is in solution or that
stream is not full.
sample stream is full (refer to Section
2.0 for installation details).
2.0 for installation details).
Open glass electrode.
Replace glass electrode.
Plugged liquid junction.
Replace liquid junction.
Recharge reference cell.
Recharge reference cell.
Reference element open-
Replace reference element.
no contact.
Meter reads between 3 and 6 pH
Electrode cracked.
Replace electrode.
regardless of actual pH of
solution or sample.
solution or sample.
Meter indication swings widely
T.C. element shorted.
Check T.C. element as instructed in
in AUTO T.C. Mode.
Section 3.5 and replace body if
defective.
defective.
Span between buffers extremely
T.C. element open.
Check T.C. element as instructed in
short in AUTO T.C. mode.
Section 3.5 and replace body if
defective.
defective.
Sluggish or slow meter indication
Electrode coated.
Clean electrode as instructed in
for real changes in pH level.
Section 3.2.1.
Electrode defective.
Replace electrode.
Transmitter cannot be
Electrode coated or cracked.
Clean electrode as instructed in
standardized.
Section 3.2.1 and, if cracked,
replace electrode.
replace electrode.
Reference chamber is contaminated
Replace gel solution as
instructed in Section 3.2.2.
instructed in Section 3.2.2.
Defective preamplifier.
Check preamplifier as instructed in
Section 3.4 and replace if
defective.
Section 3.4 and replace if
defective.
If trouble still persists, micro-junction
Replace micro-junction electrode.
electrode is defective.
Meter short spans.
Old glass electrode (greater than 1
Replace electrode.
year use).
Coated glass.
Clean with soft cloth and clean water.
TABLE 3-1. Troubleshooting
MODEL 381pH/ORP
SECTION 3.0
MAINTENANCE