Agilent Technologies E4445A User Manual

Page of 664
Chapter 4
209
Instrument Functions: P - Z
Peak Search
In
st
ru
m
e
nt
Funct
ions
:
P
-
Z
4.1
Peak Search
Places a marker on the highest peak and displays the search menu. If
Peak Search Type
(Param) is set to Excursion & Threshold, the peak found must meet the defined peak
excursion and threshold values. (See
.) Peaks that are less than 1%
of the current span away from 0 Hz are ignored. For example, if
Span
is 1 MHz, peaks will
not be found between
−10 kHz and +10 kHz. If no valid peak is found, an error (No Peak
Found
) is displayed. To clear this message, press
ESC
before attempting another search.
The peak search parameters are Peak Threshold and Peak Excursion. All searches except
Peak Search (Next Peak, Next Left, Next Right, Peak Table, SCPI “PEAKS” command)
obey the Search Parameters, which means that only peaks which rise above the Peak
Threshold by at least the Peak Excursion are found. Peak Search obeys then when Peak
Search is in PARAM mode. When the Peak Search key is in MAX mode, the search
parameters are ignored for a Peak Search.
NOTE
You can go into the Peak Search menu without actually performing a
Peak
Search
by using the front-panel
Return
key (assuming you have previously
accessed the Peak Search menu). Press
Return
to navigate through the
previously accessed menus until you return to the Peak Search menu.
Remote Command:
:CALCulate:MARKer[1]|2|3|4:MAXimum
Remote Command Notes: The :CALC:MARK:PEAK:SEARC:MODE MAX|PAR
command specifies how a peak is identified for use with the marker
commands. See
.
Example:
CALC:MARK2:MAX
performs a peak search using marker 2.
CALC:MARK2:Y?
queries the marker amplitude (Y-axis) value for marker 2.
For more information on this command, see
CALC:MARK2:X?
queries the marker frequency or time (X-axis) value for
marker 2. For more information on this command, see
.
4.1.1
Next Peak
Places the marker on the next highest peak with an amplitude less than the current peak.
The peak must meet the defined peak excursion and threshold values. Peaks that are less
than 1% of the current span away from 0 Hz are ignored. If no valid peak is found, an error
(No Peak Found) is displayed. Press
ESC
to clear this message before attempting another
search. (Also see the
and
key descriptions.)
Key Path:
Peak Search
State Saved:
Not part of saved state.