Philips V200 User Manual

Page of 328
Chapter 10
Care and Maintenance
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
10-7
Periodic 
Maintenance
This section includes detailed operator maintenance.
Schedule for Periodic Maintenance
Frequency
Component
Maintenance
During ventilator 
setup 
• Inspiratory bacteria filter
• Ventilator and patient circuit 
components
• Inspiratory and expiratory filters
• Check filter for occlusions, cracks and 
tears. 
• Perform SST (Short Self-Test) 
whenever circuit components are 
changed
• Perform EST (Extended Self-Test) 
between patient uses
• Ensure that the ventilator functions 
normally with both filters in place
At least daily, and as 
recommended by 
filter manufacturers
• Inspiratory and expiratory filters
• Monitor performance of filters and 
replace as needed. Review ventilator 
patient graphics frequently for changes 
in expiratory resistance which may 
indicate degradation of expiratory filter.
• Follow filter manufacturer 
recommendations regarding duration 
of use, maintenance (for reusable 
filters), removal and disposal. Note 
that high humidity and aerosol 
medications may reduce expiratory 
filter life, increase expiratory 
resistance, and/or cause filter damage.
CAUTION:   Do not operate the ventilator without a properly functioning 
expiratory filter and heater. Doing so may cause damage to delicate ventilator 
components, such as the expiratory flow sensor, which may lead to inaccurate 
spirometry or a Vent Inop condition.
WARNING: Vent Inop is a serious condition which is indicated by both visual and 
audible alarms. If the ventilator is attached to a patient when Vent Inop occurs, the 
patient must be supported with another means of life support ventilation.
At least daily
• Oxygen supply water trap and 
filter
• Check and empty as required every 
shift
At least every 250 
hours
• Air Inlet & Fan Filters
• Inspect and clean. Some 
environments cause a quicker 
collection of lint and dust than others, 
requiring maintenance more frequently 
than every 250 hours.
Annually
• Annual preventive 
maintenance kit (P/N 
1034840). Kit contents are 
subject to change.
• Install annual preventive maintenance 
kit.   
• Clean ventilator interior and exterior
• Complete performance verification 
procedure
CAUTION: The annual preventive maintenance procedure is to be performed only 
by a qualified service technician