Chapter 5
Setup
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
5-5
Turning the Ventilator On
1. Ensure that the Mains Circuit Breaker (refer to Figure 5-1), located on 
the ventilator back panel, is in the ON (
) position.
2. For 120V applications, if a humidifier is attached to the ventilator, 
ensure that the Humidifier AC Circuit Breaker (refer to Figure 5-1)
located on the ventilator back panel, is in the ON (
) position. A 
humidifier connection is only available on 100-120VAC ventilators.
3. Press the Power ON/OFF switch to the ON (
) position (left side up).
Turning the Ventilator Off
1. Press the Power ON/OFF Switch to the OFF (
) position (right side 
up).
Entering Diagnostic 
Mode
The Diagnostic Mode is used for:
1. Running tests of the operation of the ventilator that can only be run 
when a patient is not attached to the machine.
2. Setting altitude, time and date, and circuit compliance.
3. Other more detailed service/maintenance functions.
4. Calibration of inline Oxygen Sensor, during EST only.
Some of these functions are discussed in detail in Chapter 12, “Technical 
Specifications”.
 The functions of the Diagnostic Mode that are used to make 
the machine ready for use as a ventilator when it is first put into service are 
explained here.