Appendix B: Performance Verification (AWG710B)
AWG710&AWG710B Arbitrary Waveform Generator User Manual
B-7
Prerequisites
The tests in this section comprise an extensive, valid confirmation of performance 
and functionality when the following requirements are met:
„
The cabinet must be installed on the instrument.
„
You must have performed and passed the procedures under Self Tests, found 
on page B-2.
„
The waveform generator must have been recently adjusted at an ambient 
temperature between +20
°
 C and +30
°
 C, must have been operating for a 
warm–up period of at least 20 minutes, and must be operating at an ambient 
temperature between +10
°
 C and +40
°
 C.
Refer to Conventions on page B-1 for more information.
Equipment Required
These procedures use external, traceable signal sources to directly check warranted 
characteristics. Table B-2 lists the required equipment used to complete the 
performance tests.
Table B-2: Test equipment
Item number and 
description
Minimum requirements
Example (recommended)
Purpose
1.
Frequency Counter
(for AWG710B)
1 MHz to 3 GHz, 
Accuracy: 
< 0.2 ppm
Anritsu MF1603A
Checks reference input.
VCO Out test
1.
Frequency Counter
(for AWG710)
1 MHz to 100 MHz, 
Accuracy: 
< 0.2 ppm
Anritsu MF1603A
Checks reference input.
VCO Out test
2.
Digital multi meter
DC volts range: 0.05 V to 10 V,
Accuracy: 
±0.1 %
Fluke 8842A
Measures voltage. Used in 
multiple procedures.
3.
Oscilloscope
Bandwidth: 
> 500 MHz, 1 M
Ω
 and 
50 
Ω
 inputs
Tektronix TDS784D
Checks output signals. Used in 
multiple procedures.
4.
Oscilloscope
Bandwidth: 
> 6 GHz, 
Rise time: 
< 58.3 ps, 50 
Ω
 input
Tektronix TDS820
Checks pulse response.
5.
Function Generator
Output voltage: -5 V to +5 V, 
Frequency accuracy: 
< 0.01 %
Tektronix AFG310
Generates external input 
signals. Used in multiple input 
signal test procedures.
5.
Signal Generator
(for AWG710B)
Output voltage: -5 V to +5 V, 
Frequency accuracy: 
< 0.01 %
Agilent 8648B
Generates external input 
signals. Used in multiple input 
signal test procedures.
6.
SMA Coaxial Cable
(4 required:for AWG710B)
50 
Ω
, male to male SMA connectors
Tektronix part number 
174–1427–00
Signal interconnection
7.
BNC Coaxial Cable
(3 required)
50 
Ω
, male to male BNC connectors
Tektronix part number 
012–0482–00
Signal interconnection
8.
Adapter
(2 required)
SMA (male) to BNC (female), 50 
Ω
Tektronix part number 
015–0554–00
Signal interconnection
9.
Adapter
SMA (female) to BNC (male), 50 
Ω
Tektronix part number 
015–0572–00
Signal interconnection