Emerson I-0021 User Manual

Page of 238
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 19
Notes:
♦ 
The signal has to be stable before starting any calibration !
♦ 
If you want to start the zero and the span calibration separately, or
if you want to calibrate separately each channel of an MLT analyzer/analyzer module, or
if you want to calibrate the measurement ranges separately with span gas,
you have to change to the menu "Analyzer Module Calibration" (see p. 5-15) resp. to
the menu "Calibration Procedure Setup" (p.5-9).
♦ 
You can cancel any calibration procedures with the function "Cancel all running
procedures !"
♦ 
Via the F2 key you will reach the menu "Analyzer Channel Status". In that menu and
its corresponding submenus you will find the status report of the current channel for
• 
Failures
• 
Maintenance requests
• 
Function controls
• 
Measurements/Alarms
• 
Events
• 
Hours of operation
and the operational settings of the current channel like
• 
range settings and
• 
response time (t
90
-time).
The menu "Analyzer Channel Status" and its submenus are displayed detailed in
section 4.1.
Menu line "Info...":
Via this menu line you can change to the submenu "State of Calibration Procedures",
where you will find the actual status of each channel of an MLT analyzer or analyzer
module.
Measure
Back...
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- State of Calibration Procedures --
Channel -1:
Ready
Channel -2:
Ready
Channel -4:
Ready
Channel -5:
Ready
Channel -3:
Ready