Emerson I-0021 User Manual

Page of 238
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 59
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Gasflow setup
↓↓
In the menu "Gasflow Setup" you can set up the modus of gas flow through
the analyzer cell:
♦ 
Parallel or
♦ 
Serial
and the valve position at "Basic-Status":
♦ 
Samplegas or
♦ 
All closed
Notes:
♦ 
The adjustment for the gas flow through the analyzer cell depends on the real
combination of the analyzer cells.
♦ 
The setup is normally correct by factory setting.
You have only to change it, if you modify your configuration of cells.
♦ 
The correctness of this adjustment is very important for the time controlled calibration
with valve gear (see 5.1.1 p. 5-12)
Set up parameters:
♦ 
Change to the line you want with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Press the 
88 
 -key or the 
 -key to select the variable.
♦ 
Select the parameter you want with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Confirm your adjustment with the 
88 
 -key or
cancel and go back to the last one with the F2 key.
Measure
Back...
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Gasflow Setup --
Gasflow through analyzer cell is:
Parallel
Valve position at "Basic-Status":
Samplegas
Channel
Operation status:
Ready
Valve position:
Samplegas
5.1 Analyzer Module Setup
5.1.13 Gasflow Setup
[2nd menu page via F5 (More...)]