Emerson I-0021 User Manual
4 - 48
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Yes
Back...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
TAG
F1
F2
F3
F4
F5
37.50 ppm
- S U C C E S S -
- The selected function has been started/executed -
(Wait a moment...)
Measure
Status...
Channel
Back...
TAG
F1
F2
F3
F4
F5
0.00 ppm
-- Set Gas Valves --
Flow sample gas !
Flow zero gas !
Valve position:
Zerogas
Operation status:
Ready
Flow span gas !
Close all valves !
Flow test gas !
⇒
⇒
Confirm the order to
flow zero gas
flow zero gas
Press
F2
to start the zero
gas flow immediately.
gas flow immediately.
Option:
Press the F4 key if you want
to cancel the order and go
back to the menu "Set Gas
Valves".
Press the F4 key if you want
to cancel the order and go
back to the menu "Set Gas
Valves".
Note:
The display of this message
depends on the setup in the
expert configurations.
(see 5.1.8 p. 5-49)
The display of this message
depends on the setup in the
expert configurations.
(see 5.1.8 p. 5-49)
⇒
⇒
Confirmation
message of the
function start
appears
message of the
function start
appears
Note:
This message will be
displayed after the function
has been started. Then the
display will jump automatic-
ally to the menu "Set Gas
Valves".
This message will be
displayed after the function
has been started. Then the
display will jump automatic-
ally to the menu "Set Gas
Valves".
⇒
⇒
Change to the single
component display
of the current
channel
component display
of the current
channel
Options:
•
Start the gasflow with
span gas, sample gas
or test gas.
span gas, sample gas
or test gas.
•
Close all valves.
•
F3 key: Changing to
another available channel
to execute gas flow.
another available channel
to execute gas flow.
•
F4 key: Go back to the
menu "Analyzer Module
Calibration" to start
zeroing or spanning.
menu "Analyzer Module
Calibration" to start
zeroing or spanning.