Chapter 3 Operation
 
IM-83/84/86-H 
3-1
 
Chapter 3  Operation 
This chapter covers transducer operation and controller features 
available on Sierra’s Side-Trak
 and Auto-Trak
 models. For 
more information on the advanced digital RS-485 commands 
available on Auto-Trak transducers, refer to the Series 860 “Auto-
Net” Software Addendum included with this manual.  
 
Standard output for all transducers is a linear 0-5 VDC output signal 
directly corresponding to 0 to 100% of the mass flow full scale 
range. For mass flow controllers an input signal of 0-5 VDC (4-20 
mA optional) allows set point flow control to any desired value 
within the range of the model. The input signal is a direct linear rep-
resentation of 0 to 100% of the mass flow full scale value. A 0 VDC 
(or 4 mA) set point will cause a condition of 0% flow to occur and a 
5.00 VDC (20 mA) set point will cause a flow condition equivalent 
to 100% of flow to occur. 
Mass Flow Meter Operation 
When the transducer is installed and the system has undergone a 
complete leak check: 
 
1.  Apply power. Allow thirty minutes of warm-up time. (When 
power is first applied, the output signal from the transducer 
remains fixed at a much higher than normal level until the sen-
sor warms up to its normal operating temperature range.) 
 
2.  Turn on the gas supply. Perform an initial zero output check 
(only required for first-time start ups). Set flow to zero. Connect 
a digital multimeter to V+out (4-20 mA out) and COM termi-
nals. 
 
3.  Check the reading. If it is not within ±10 mV (±0.10 mA) of 
zero, adjust the zero potentiometer to zero the transducer. (The 
zero pot is located behind the upper metal button on the side of 
the transducer or behind the swing-out plastic door.)