P/N 06160D DCN6409 Teledyne API Model 465M O
3
 Monitor Instruction Manual - Page 66 
 
10.2. 
Troubleshooting Using Front Panel Status 
LED’s or Status Outputs 
The M465M has 4 front-panel status LED’s that reflect the current operating status of the 
monitor, and indicate fault conditions.  There are also four relay Status Outputs on the 
rear panel (or internally in the NEMA configuration) that also reflect the state of these 
status LED’s.  See Section 4.5.3 for more information on the Status Outputs. 
Table 10-1:  Status LED/Output Definitions 
Status 
Output # 
Status LED 
Label 
Normal 
Operating 
State 
Triggers 
Critical 
Warning? 
1 Sensor 
OK 
On 
  Reference < 125.0 mV 
  Reference > 1230.0 mV – 0.5 Wt% 
Yes 
2 Invalid 
Reading 
Off 
  O3 Concentration < F.S. Range 
  O3 Concentration > F.S. Range 
No 
3 Check 
Lamp 
Off 
  Reference < 250.0 mV 
No 
4 Pneumatic 
Error 
Off 
  Pressure < 9.0 psia 
  Pressure > 18.0 psia 
  Flow < 500 cc/min* 
No 
*If Flow Switch installed