Emerson 54e User Manual
3
MODEL 54e pH/ORP
SPECIFICATIONS
PHYSICAL SPECIFICATIONS – GENERAL
Enclosure: Epoxy-painted aluminum
NEMA 4X (IP65)
144 X 144 X 131 mm, DIN size (5.7 X 5.7 X 5.2 in.)
144 X 144 X 131 mm, DIN size (5.7 X 5.7 X 5.2 in.)
Front Panel: Membrane keyboard with tactile
feedback and user selectable security. Light gray,
blue and white overlay
Light gray enclosure, dark gray bezel
blue and white overlay
Light gray enclosure, dark gray bezel
Display: Back-lit dot matrix LCD (7.0 x 3.5 cm),
blue on gray-green
The display contrast is compensated for ambient
temperature.
The display contrast is compensated for ambient
temperature.
Process Variable Character Height: 16mm (0.6 in.)
Electrical Classification:
Class I, Division 2, Groups A, B, C, & D.
T5 Ta=50°C. Dust ignition proof: Class II,
Division 1, Groups E, F, & G; Class III.
T5 Ta=50°C. Dust ignition proof: Class II,
Division 1, Groups E, F, & G; Class III.
CSA-LR34186:
Max. relay contact rating: 28 Vdc; 110 Vac;
230 Vac; 6 amps resistive
230 Vac; 6 amps resistive
FM: Max. relay contact rating: 28 Vdc resistive
150 mA - Groups A & B;
400 mA - Group C;
540 mA - Group D
400 mA - Group C;
540 mA - Group D
Power:
Code -01: 100 - 127 VAC, 50/60 Hz ± 6%, 6.0 W
200 - 253 VAC, 50/60 Hz ± 6%, 6.0 W
Code -02: 20 - 30 VDC, 6.0 W
Current Outputs: Output 1: pH, ORP, temperature,
glass impedance, or reference impedance.
Output 2: pH, ORP, temperature, glass impedance,
or reference impedance.
or reference impedance.
Each output is galvanically isolated, 0-20 mA or
4-20 mA into 500 ohms maximum load at 24 Vdc
or 115/230 Vac or 500 ohms maximum load at
100/200 Vac. Output 1 includes digital signal
4-20 mA superimposed HART (Code -09 only).
4-20 mA into 500 ohms maximum load at 24 Vdc
or 115/230 Vac or 500 ohms maximum load at
100/200 Vac. Output 1 includes digital signal
4-20 mA superimposed HART (Code -09 only).
RFI/EMI: EN-61326
LVD (Code -01 only): EN-61010-1
Ambient Temperature: 0 to 50°C (32 to 122°F)
NOTE: The analyzer is operable from -20 to 60°C
(-4 to 140°F) performance.
(-4 to 140°F) performance.
Relative Humidity: 95%, non condensing
Alarms:
Relay 1 - Process, Interval, or Time Proportional
Control (TPC requires code -20)
Relay 2 - Process, Interval, or Time Proportional
Control (TPC requires code -20)
Control (TPC requires code -20)
Relay 3 - Process, Interval, or Time Proportional
Control (TPC requires code -20)
Control (TPC requires code -20)
Relay 4 - Sensor/analyzer and process fault alarm
Each relay has a dedicated LED on the front
panel.
panel.
Relay Contacts: Relays 1-3: Epoxy sealed form A
contacts, SPST, normally open
Relay 4: Epoxy sealed form C, SPDT
Resistive
Inductive
28 Vdc
5.0 Amps
3.0 Amps
115 Vac
5.0 Amps
3.0 Amps
230 Vac
5.0 Amps
1.5 Amps
Weight/Shipping Weight: 1.8 kg/2.3 kg (4 lb/5 lb)
INSTRUMENT SPECIFICATIONS @ 25°C
— pH
— pH
Measurement Range: 0 to 14 pH
Output Scale Expansion: Zero suppression: up to 13
pH units
Span: Any pH from 1 to 14
Span: Any pH from 1 to 14
Accuracy: ± 0.01 pH
Repeatability: ± 0.01 pH
Stability: ± 0.01 pH/month, non-cumulative
Temperature Coefficient: Input: ± 0.003 pH/°C
Output: ± 0.006 pH/°C
Temperature Compensation: Pt 100 or Pt 1000 RTD,
Automatic or Manual; –15 to 120°C (5 to 248°F)
INSTRUMENT SPECIFICATIONS @ 25°C
— ORP
— ORP
Measurement Range: –1400 to +1400 mV
Output Scale Expansion: Zero suppression: up to
±1300 mV
Span: Any ORP range from 100 to 2800 mV
Span: Any ORP range from 100 to 2800 mV
Accuracy: ± 1.0 mV
Repeatability: ± 1.0 mV
Stability: ± 1.0 mV/month, non-cumulative
Temperature Coefficient: Input: ± 0.2 mV/°C
Output: ± 0.4 mV/°C
Temperature Measurement: –15 to 120°C (5 to 248°F)
Pt 100 or Pt 1000 RTD
Sira MC070111/00