A&D HV-G Series User Manual

Page of 69
HV-G Instruction Manual Page 41
14.2 The Complete Calibration Procedure
14.2.1 Gravity Acceleration Correction
Step1
Turn on the display.
Open the rear cover of the display unit.
Inside the display unit, press and hold the
CAL 
 switch to enter the calibration mode.
Then 
Cal 0 
 is displayed.
Step2
Press the 
 ZERO 
 switch to enter the gravity
acceleration correction mode.
Step3
Set your local gravity acceleration using
the following switches.
^
 
 switch Selecting the value of a digit.
<
 
 switch Selecting a digit.
Step4
Press the 
 ENTER 
 switch to store the new
value.
Step5
Press the  
CAL
  switch again. Then the
scale returns to the normal weighing
mode.
14.2.2 Preparation
Step 6 Confirm the environmental conditions as follows:
Maintain a constant temperature and a stable power supply.
Install the scale on a solid floor or bench where there is no draft,
vibration, strong magnetic fields or direct sunlight.
Consider section “4. Caution”.
Step 7 Display normal weighing for at least 30 minutes to warm up the scale.