A&D HV-G Series User Manual

Page of 69
HV-G Instruction Manual Page 48
16 RS-232C Serial Data Interface
RS-232C interface has the following two modes.
Stream mode
Outputs data countinuously and can be used for printing data.
Command  mode Controls the scale using commands from a computer.
It is necessary to set the parameters of the “Baud rate ( 
 f4 
 )”,
”Output mode ( 
 f5 
 )”, “ Format ( 
 f15 
 )” in the function table in advance.
There are option cables as follows:
AX-KO557A RS-232C cable, D-sub 25 pin, 2m
AX-KO1786-200 RS-232C cable, D-sub 9 pin, 2m
Transmission system
EIA RS-232C
Transmission
Asynchronous, bi-directional, half-duplex
Data format
Baud rate: 2400, 4800, 9600 bps
Data:
7 bits
Parity:
1 bit, Even
Start bit
1 bit
Stop bit
1 bit
Code
ASCII
Terminator
CR LF   (CR: 0D[H], LF: 0A[H])
Pin Connections
Pin No. Signal name
Direction Description
1
-
2
RXD
Input
Receive data
3
TXD
Output
Transmit data
4
-
5
SG
-
Signal ground
6
-
7
DSR
Output
Data set ready
Circuit