16.2 Stream Mode
The scale outputs the current display data at the same time as refreshing the display.
The scale does not output data while in the setting mode.
Averaging  function table
Refresh rate
f13 0
Normal weighing
Approximately 7 times/s while the display is
 unstable.
Approximately 4 times/s when the display is stable
f13 1
A person’s weight
Approximately 4 times/s
f13 1
Animal weighing
Approximately 4 times/s
Step 1
Turn off the display.
Press the  
ON/OFF
  switch while the  
ZERO 
switch is
pressed and held. The function table is displayed.
Step 2
Select a parameter for the “baud rate ( 
 f4 
 )”
using the 
 ENTER 
 and 
 
^
 
 switches.
Step 3
Select “stream mode ( 
 f5 0 
 )” of “output” using
the 
 ENTER 
 and 
 
^
 
 switches.
Step 4
Select “format 1 ( 
 f15 0 
 )” of “format” using the
ENTER 
 and 
 
^
 
 switches.
Step 5
Press the  
ENTER
  switch to store the new
parameters. The scale returns to normal weighing
and outputs the data countinuously.
16.2.1 Preparation and Performance (Examples)