Counting operation 
7  Place the samples to be counted on the pan.
 
Note 
  Up to 50 unit masses can be stored in memory for the multiple sample. For details, refer to   
“11. DATA MEMORY”. 
 
Counting mode using the ACAI function 
The ACAI is a function that improves the accuracy of the unit 
mass automatically by increasing the number of samples as 
the counting process proceeds. 
8  If a few more samples are added, the processing indicator turns on. To prevent an error, add 
three or more. The processing indicator does not turn on if overloaded. Try to add the same 
number of samples as displayed. 
9  The balance re-calculates the unit mass while the processing indicator is blinking. Do not 
touch the balance or samples on the pan until the processing indicator turns off. 
10  Counting accuracy is improved when the processing indicator turns off.   
 
Each time the above operation is performed, a more accurate unit mass will be obtained. 
There is no definite upper limit of ACAI range for the number of samples exceeding 100. Try 
to add the same number of samples as displayed. 
11  Remove all the samples used in ACAI and proceed with the counting operation using the 
improved unit mass.
 
Note 
ACAI will not function on the unit mass entered using the keys, or digital input mode.