A&D GP-60K User Manual

Page of 98
 
 73
12. PROGRAMMABLE-UNIT 
This is a programmable unit conversion function. It multiplies the weighing data in kilograms by an 
arbitrary coefficient set in the function table and displays the result.   
The coefficient must be within the range between the minimum and maximum shown below. If the 
coefficient set is beyond the range, an error is displayed and the balance returns to the coefficient 
setting mode, prompting to enter an appropriate value. A coefficient of 1 was set at the factory. 
Model 
Minimum coefficient Maximum coefficient 
GP-12K/20K/30K/30KS/40K/61K/61KS
10000 
GP-60K/60KS/100K/100KS 1000 
GP-32K/32KS 1000 
GP-102K 
 
0.000001 
100 
 
 
Operation 
1  Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the 
key. 
2  Press the SAMPLE key several times to display    mlt
Cadj
 . 
3  Press the PRINT key. The balance enters the mode to confirm or set the coefficient. 
Confirming the coefficient 
4  The current coefficient is displayed with the first digit blinking.   
  
When it is not to be changed, press the CAL key and proceed to step 6. 
  
When it is to be changed, press the RE-ZERO key 
and proceed to step 5. 
Setting the coefficient 
5  Set the coefficient using the following keys. 
SAMPLE key 
To select a digit to change the 
value. The selected digit blinks. 
RE-ZERO key 
To change the value. 
MODE key 
To change the decimal point 
position. 
 
Each time the switch is pressed, 
the decimal point position 
changes as follows: 
 
 
 
PRINT key 
To store the new setting, display end and go to step 6. 
CAL key 
To cancel the new setting and go to step 6. 
Quitting the operation 
6  The balance displays    Unit . Press the CAL key to exit the programmable-unit function and 
return to the weighing mode. 
Using the function 
Press the MODE key to select the programmable-unit (no display on the unit section). Perform 
weighing as described in “5-1 Basic Operation (Kilogram Mode)”. After weighing, the balance 
displays the result (weighing data in kilograms x coefficient).