Chapter 10    Peripherals    
107
10
Peripherals
Fitting the Combined Interface Module
To fit the Combined Interface Module:
    
          
 
1. Remove the screw which holds the interface socket cover in place on the base of the 
monitor and then remove the interface socket cover.
2. Connect the Combined Interface Module.
3. Screw the module in place.
Figure 10-1Fitting the Combined Interface Module
With the module fitted, you can connect the following peripherals to the monitor:
l
One of the following to the RS232 serial interface:
n
HBCR8200 Barcode Reader.
n
Dinamap
 1846/8100 maternal NIBP Monitor.
n
COLIN Press-Mate
/Nippon Colin Model BP-8800 maternal NIBP Monitor.
n
Nellcor OxiFirst
 Fetal Oxygen Saturation monitor.
l
M1310A Fetal Telemetry to the telemetry socket.
l
M1383A/B/C OB TraceVue to the system interface.