Drawer Assembly
150
Chapter 11    Replacing Parts    
Drawer Assembly  
To remove the drawer assembly:
Figure 11-13 Removing the Paper Table
1. Remove the top cover and the Recorder Board. 
2. Place the cover with the paper table facing upwards and slide back the drawer 
assembly. 
3. Remove the two countersunk screws (S2). 
4. Turn over the cover and place it so that the stepper motor is on the left and facing 
upwards.
5.  Remove the RFI bracket: 
a.
Remove the two screws (S1 - one on left side, one on right side). 
b.
Pull up and lift out the RFI bracket (it is fitted with a lug). 
6. Remove the recorder chassis:
a.
Remove the two screws (S3 - one on left side, one on right side). 
b.
Lift out the recorder chassis.
7. Slide the paper table through the gap exposed by the recorder chassis.
To refit the paper table, reverse the above procedure taking note of the following:
Secure the two countersunk screws (accessed from top cover) before securing the 
recorder-chassis screws. 
After refitting the paper table, carry out the System Test (see page 105).