Philips M1351A User Manual

Page of 189
Chapter 2    Technical Specifications    
13
2
Technical Specifications
Monitor
Power 
Requirements
The monitor is set for the correct voltage at the factory. Before you connect power, however, 
ensure that the voltage label shows the correct setting for your country.
        
   
Operating Voltage:
100V - 120V
 or 220V - 240V (±10%).
Line Frequency:
50 to 60Hz ±5%.
Power Consumption:
25VA max.
Battery Type:
2 x 1.5V (AA size). Lifetime > 1 year.
Environment
The monitor should be used in an environment which is reasonably free from vibration, 
dust, corrosive or explosive gases, extremes of temperature, humidity, etc. It operates within 
specifications at ambient temperatures between 0 and 55°C. Ambient temperatures which 
exceed these limits can affect the accuracy of the monitor and cause damage to the 
components and circuits. Allow at least 5cm (2in) clearance around the monitor for proper 
air circulation.  
       
Operating Temp:
0 to +55°C (32°F to 131°F).
Storage Temp:
-40 to +75°C (-40°F to 167°F), excludes transducers: -40 to 
+60°C (-40°F to +140°F)
Relative Humidity:
5 to 95%.
Weight and 
Dimensions
Height:
115mm (4.5in). 
      
Width:
340mm (13.4in).
Depth:
308mm (12.1in). 
Weight:
5.74kg (12.6lb) (without transducers).