Philips M1351A User Manual
Configuring the Monitor Using a PC
32
Chapter 4 Configuring the Monitor
Main Menu
To select an item from the menu, move the cursor to the item you require and press
Enter
.
Use the
,
,
,
keys on the keypad, or the
E
,
S
,
D
or
X
keys to move the cursor.
While the service program is working, messages maybe displayed along the bottom of the
screen. For example:
screen. For example:
Reading Recorder Adjustments from the Monitor.
Reading Serial Number from the Monitor.
Clearing the Error Log.
Configuring the
Monitor
To change a setting:
Series 50 Fetal Monitor Configuration and Service Software Rev A.nn.nn
Main Menu
Configuration Settings
Recorder Adjustments
Service Menu
Exit
MESSAGE:
Select with >cursor keys< or >E, S, D, X< then press <ENTER> to execute!!
Series 50 Fetal Monitor Configuration and Service Software Rev A.nn.nn
Configuration Setting:
Function
:
Actual Setting is
Select
Paper Scale :
US
US EU
Time Format :
AM/PM
AM/PM 24h
Date Format :
US
US EU
IUP Scale
:
mmHg
mmHG kPA
MESSAGE:
Select with >cursor keys< or >E, S, D, X< then press <ENTER> to execute!!