Tests and Error Messages
Chapter 7    Tests and Error Messages    
69
Operator Error Messages 
Table 7-2 Operator Error Messages
Message
Display
Cause and Solution
Err 1
1
1. See the Instructions for Use for your monitor for the types of transducers that can be connected.
US/US1
Wrong transducer in the US1/US socket. 
Connect the correct transducer.
Err 1
1
US2/ECG
Wrong transducer in the US2/ECG socket.
Connect the correct transducer.
Err 2
1
 
Toco 
Wrong Toco transducer in the Toco socket.
Connect the correct transducer.
Err 4
1
US
ECG/US2 
US/MECG Combi transducer not allowed.
Remove the transducer.
Err 9
 US1/US 
Toco 
US2/ECG 
Invalid Telemetry mode.
Check the cable from the Telemetry Receiver and, if necessary, replace it. Check all connections.
Err 16
US1/US 
Toco 
US2/ECG
Wrong pairing of Telemetry and transducers.
Either disconnect the transducers or switch off the Telemetry Receiver.
nop
 
ECG
No contact or poor contact between any of the contacts of the system. Check all connections, espe-
cially the legplate connector electrode.
Use a new spiral electrode.
Err PAP
2
/
Err 18
2. Displayed for ten seconds after the wrong paper is detected.
 
Incorrect type of paper loaded. 
Replace with recommended paper.
Err bAt 
3
/ Err 19
3. Displayed for ten seconds when the monitor is first switched on.
Battery low or empty of charge. 
Change the batteries as soon as possible. If you do not change the batteries, your specific settings will return 
to their default values when the monitor is switched off (for example, the date is set to 4.4.44).
Error 601 
 
5
 
Paper speed.
Check that correct paper is used.
Check the speed by timing how long it takes for the paper to advance 1cm:
60 seconds = 1cm/min
30 seconds = 2cm/min
20 seconds = 3cm/min
Contact your Philips Service Engineer or Response Center if the speed is incorrect.
Err PAP 30-240 
4
Err 602
5
 
4. Printed every ten minutes.
5. Printed every three pages.
Incorrect type of paper loaded. 
Load paper with 50-210 scale or change the paper format setting of the monitor.
Err PAP 50-210
 
4
Err 602
5
Incorrect type of paper loaded. 
Load paper with 30-240 scale or change the paper format setting of the monitor.
Err 70
6
6. Only applicable with the Modem Interface Module fitted.
Modem not responding.
The Modem is either incorrectly installed, faulty or not an approved model.
Err 77
6
Telephone transmission failure.
Line busy or faulty, wait short time and send information again.
Err 532
6
Modem Interface board failure.
Board failure will prevent data transmission from taking place and may possibly generate 
Err 70
 and 
Err 77
 messages.