dhi interconnection kit for mpg1 with rpm4 or pg 7302 pn 402265 instruction sheet User Manual
INTERCONNECTION KIT FOR MPG1 WITH
RPM4 OR PG 7302, P/N 402265, INCLUDES:
RPM4 OR PG 7302, P/N 402265, INCLUDES:
DESCRIPTION
QTY
PART NO.
DH500 6 in. nipple
1
100208
DH500 12 in. nipple
1
100209
DH500 nipple, 5 in. x 1/8 in.
2
123019
DH500 tee
1
100291
DH500 plug
1
100285
Adaptor, DH500F x 1/8 in. NPTM
2
102819
INTRODUCTION
This kit provides the tubing and fittings necessary to
connect an MPG1 pressure controller to either an
RPM4 reference pressure monitor or to a PG 7302
piston gauge, and also to a device under test. This
document contains instructions for connecting the
fittings and tubing.
connect an MPG1 pressure controller to either an
RPM4 reference pressure monitor or to a PG 7302
piston gauge, and also to a device under test. This
document contains instructions for connecting the
fittings and tubing.
INSTALLATION OF INTERCONNECTIONS
n Remove Plugs from MPG1 and RPM4 or PG
7302: Unscrew the glands from the MPG1’s test
ports and RPM4’s or PG 7302’s test port.
ports and RPM4’s or PG 7302’s test port.
Remove the collar and plug from inside each
gland. There should be two plastic plugs and one
steel plug. The plastic plugs will not be used.
Keep the steel plug.
gland. There should be two plastic plugs and one
steel plug. The plastic plugs will not be used.
Keep the steel plug.
o Install glands and collars onto flexible 5 in.
nipples: Slide a gland over a nipple so that the
gland’s threads face outward. Next, screw a collar
onto the nipple by turning it to the LEFT. Allow
two nipple threads to show between the collar and
nipple tip.
gland’s threads face outward. Next, screw a collar
onto the nipple by turning it to the LEFT. Allow
two nipple threads to show between the collar and
nipple tip.
p
a
RPM4 – Install glands and collars onto the 6 in.
nipple.
p
b
PG 7302 – Install glands and collars onto the
12 in nipple.
q Install nipples into three way tee: Install the 5
in. nipple into the center port of the three way tee.
Install either the 6 in. or 12 in. nipple into either of
the end ports on the three way tee as shown in
figure 1. Tighten the nipples with a 5/8 in. wrench
or crescent wrench. The thick nipple on the right
connects to the MPG1. The thin nipple in the
center of the plug connects to the RPM4 or to the
PG7601. In figure 1 the left port is currently
plugged, but the remaining thin 5 in. nipple can go
in the left port and connect to the DUT.
Install either the 6 in. or 12 in. nipple into either of
the end ports on the three way tee as shown in
figure 1. Tighten the nipples with a 5/8 in. wrench
or crescent wrench. The thick nipple on the right
connects to the MPG1. The thin nipple in the
center of the plug connects to the RPM4 or to the
PG7601. In figure 1 the left port is currently
plugged, but the remaining thin 5 in. nipple can go
in the left port and connect to the DUT.
Figure 1: Two nipples installed in 3-way tee.
r Optional: Extend the front feet on RPM4: The
RPM4’s front feet can be extended so that the RPM4
is inclined for easier reading and display access.
is inclined for easier reading and display access.
s RPM4 A70M or less - install NPT adapter into
RPM4. RPM4s that have ranges less than 70
MPa (10,000 psi) have 1/8 in. NPTF test ports.
These ports require the use of one of the DH500F
by 1/8 in. NPT adapters. Wrap Teflon thread tape
around the NPT adapter’s threads three times,
and install the adapter into the RPM4.
MPa (10,000 psi) have 1/8 in. NPTF test ports.
These ports require the use of one of the DH500F
by 1/8 in. NPT adapters. Wrap Teflon thread tape
around the NPT adapter’s threads three times,
and install the adapter into the RPM4.
t Install the free end of the center nipple into
RPM4 or PG 7302: The center, thin, 5 in. nipple
should be inserted into the RPM4’s or the PG 7302’s
test port. Screw in the gland in loosely by hand.
The nipple can be bent slightly to ensure that nipple
enters straight into the test port. Do not tighten the
gland in the RPM4 or PG 7302 yet, since it will be
tightened when the system is primed.
should be inserted into the RPM4’s or the PG 7302’s
test port. Screw in the gland in loosely by hand.
The nipple can be bent slightly to ensure that nipple
enters straight into the test port. Do not tighten the
gland in the RPM4 or PG 7302 yet, since it will be
tightened when the system is primed.
Interconnection Kit
For
MPG1 with RPM4 or PG 7302
P/N 402265
Instruction Sheet