7.  Rinse the probe with de-ionised water or a rinse 
solution, and place it in the next pH buffer.  
8.  Follow steps 5 and 6 for additional calibration 
points. See Figure 12. 
9.  When all the calibration points set in the Unit 
Configuration Setup (see P1.3 Select Number 
of pH Calibration Points on page 36) are 
completed, the meter returns to Measurement 
mode automatically. However, if you wish to 
terminate the calibration without completing the 
number of points as set in the Unit 
Configuration Setup menu, press CAL/MEAS to 
return to pH measurement mode.  
 
 
NOTES: 
1. To exit from pH calibration mode without confirming 
calibration, DO NOT press ENTER in step 6.  Press 
CAL/MEAS instead.  
2.  To limit the number of pH buffer values available during 
calibration, see P1.3 Select Number of pH Calibration Points 
on page 36.  
 
 
 
 
 
CAL
pH
pH
READY
4.15
4.01
CAL
pH
pH
READY
4.01
4.01
ENTER
HOLD
Figure 12: Proceed to next 
calibration point