eutech-instruments cyberscan ph 110 phmvrs232 meter User Manual

Page of 76
Instruction Manual   
 
CyberScan pH 11 / 110  
 
64
14.3  Standard pH Buffers 
The following table shows the various pH values at different temperature of the 
solution during calibration. 
Temperature 
(ºC) 
pH 1.68 
(oxalate) 
pH 4.01 
(phthalate) 
pH 7.00 
(neutral 
Phosphate) 
pH 10.01 
(carbonate) 
pH 12.45 
(Saturated 
Calcium 
Hydroxide 
Solution) 
0  1.67 4.01 7.12 10.32 13.43 
5  1.67 4.01 7.09 10.25 13.21 
10  1.67 4.00 7.06 10.18 13.00 
15  1.67 4.00 7.04 10.12 12.81 
20  1.68 4.00 7.02 10.06 12.63 
25  1.68 4.01 7.00 10.01 12.45 
30  1.69 4.01 6.99 9.97 12.29 
35  1.69 4.02 6.98 9.93 12.13 
40  1.70 4.03 6.97 9.89 11.99 
45  1.70 4.04 6.97 9.86 11.84 
50  1.71 4.06 6.97 9.83 11.70 
55 
-  4.08 6.97 9.81  - 
60 
-  4.10 6.98 9.79  - 
70 
-  4.12 6.99 9.76  - 
80 
-  4.16 7.00 9.74  - 
90 
-  4.20 7.02 9.73  -