16.2  pH 110 Factory Default Settings 
PROGRAM FUNCTION  PARAMETER 
OPTING KEYS 
PARAMETER 
OPTIONS 
DEFAULT 
SETTING 
P 1.0 
Unit Configuration 
Parameter Title 
Only
 
Parameter Title 
Only 
 
P 1.1 
Ready Indicator 
MI/▲; MR/▼ 
ON ; OFF, HOLD 
ON 
P 1.2 
Select pH buffer option 
MI/▲; MR/▼ 
USA, nSt, dn, PUr 
USA 
P 1.3  
Number of calibration 
points 
MI/▲; MR/▼ 
3 P, 4 P, 5 P, 2 P 
3 P 
P 1.4 
Select °C or °F 
MI/▲; MR/▼ C, 
P 2.0 
View previous calibration 
data 
Parameter Title 
Only
 
Parameter Title 
Only 
 
P 2.1 
First calibration point 
Viewing Only 
Viewing Only 
- - - 
P 2.2 
Second calibration point 
Viewing Only 
Viewing Only 
- - - 
P 2.3  
Third calibration point 
Viewing Only 
Viewing Only 
- - - 
P 2.4 
Fourth calibration point 
Viewing Only 
Viewing Only 
- - - 
P 2.5 
Fifth calibration point 
Viewing Only 
Viewing Only 
- - - 
P 2.6 
Sixth calibration point 
Viewing Only 
Viewing Only 
- - - 
P 3.0 
View electrode data 
Parameter Title 
Only
 
Parameter Title 
Only 
 
P 3.1 
Electrode offset (pH mode)  Viewing only 
Viewing only 
- - - 
P 3.1 
Electrode offset (mV 
mode) 
Viewing only 
Viewing only 
0.0  
P 3.2 
pH electrode slope 
Viewing only 
Viewing only 
- - - 
P 4.0 
Select auto power on/off 
MI/▲; MR/▼ YES, 
nO 
YES 
P 5.0 
Reset to factory default 
Parameter Title 
Only
 
Parameter Title 
Only
 
 
P 5.1  
Reset pH or ORP 
calibrated data 
MI/▲; MR/▼ nO, 
YES 
nO 
P 5.2 
User reset to factory 
default 
MI/▲; MR/▼ nO, 
YES 
nO 
P 6.0 
Memory clear 
MI/▲; MR/▼ nO, 
YES 
nO 
NOTE: The HOLD/ENTER key is used to confirm every parameter opted. The 
CAL/MEAS on the other hand will escape the set up operation to return to the 
measurement mode.