 • If the reading is not close
to the selected buffer, the
"BUF" prompt will continue
to blink.
If on the other hand, the
reading is close to the se-
lected buffer, the meter
will recognize the standard
solution. The BUF symbol
will stay still on the dis-
play, while the "
" sym-
bol will blink, indicating that
the meter is stabilizing its
reading.
 • Once stabilized "CON"
symbol will appear on
the display.
Wait for a couple of min-
utes and press the CFM
key.
 • The meter will return to its normal mode
and the two point calibration process is
completed.
CFM
mannacl_r1.P65
18
18
NaCl CALIBRATION (HI931100/HI931102)
The calibration is performed by immersing
the FC 300B sodium-sensitive electrode in 2
different standard NaCl solutions to achieve
accurate 2 point calibration.
It is also possible to calibrate just the offset
(3.00 g/L NaCl) following the single point cali-
bration procedure described below.
For best accuracy, a two-point calibration is
recommended (see below).
PREPARATION
  •
For HI931100 and HI931102:
Pour small quantities of
HI 7083 or HI8083 (3.00 g/L
NaCl) solution and HI7085
or  HI 8085 (0.30 g/L) or
HI 7081 or HI 8081 (30 g/L)
into two clean beakers.
To get accurate readings use HI7083/HI8083
and HI 7085/HI8085 if you are going to mea-
sure samples with low sodium chloride con-
tent (below 3 g/L) or HI7083/HI8083 and
HI7081/HI8081 if you are going to measure
samples with high sodium chloride content
(over 3 g/L).
  • For accurate calibra-
tion, use two bea-
kers for each buffer
calibration, the first
one for rinsing the electrode, the second
one for calibration. This way, contamina-
tion of the buffer is mini-
mized.  Whenever possible
use plastic beakers to mini-
mize any EMC interferences.