fisher--paykel mr850 User Manual

Page of 62
 
MR850 Technical Manual   *   Revision J   * Issued November 2005   *   Ref.  185041340 (185041713 internal)
 
20 
 
Figure 4.3 Temperature vs Time to Alarm 
Pressing mute during a temperature alarm silences the alarm for half the normal 
time period, if the same temperature is maintained. 
The low temperature warning and alarm can be caused by cold or drafty 
ambient conditions, or can result from using gas flow rates outside the 
specification of the breathing circuit, chamber or humidifier. 
NOTE: The low temperature alarm is disabled in stand-by on software version 
5.45 and 5.70. 
4.5.2 
Non-Heater Wire Operation (Software versions 7.22 & 7.23) 
 
When the airway temperature exceeds 41 °C, the heater plate is shut off, and 
an immediate visible and audible high temperature alarm is activated. 
In invasive mode, the low temperature warning and alarm is identical for the 
heater wire mode (see above) after warm-up. 
In addition the airway temperature must reach 29.5 °C in invasive mode (or 26 
°C for non-invasive mode) within 15 minutes, otherwise a visible and audible low 
temperature alarm will be given, and the heater plate will be switched off.  This 
will occur if the airway probe has not been inserted into the breathing circuit. 
After warm-up, if the airway temperature drops below 29.5 °C (or 26 °C non-
invasive mode), a visible and audible low temperature alarm will be given, and 
the heater plate will be switched off. 
NOTE: The low temperature alarm will be inactive in stand-by. 
4.5.3 See 
Manual 
Indicator 
 
This indicates a serious hardware fault.  Please refer to section 6, 
"Troubleshooting".