Hanna Instruments hi 964400 User Manual

Page of 48
6
Range
D.O.  ppm
0.00 to 45.00
D.O.  %
0.0 to 300.0
Temp. °C
0.0 to 50.0
Resolution D.O. ppm
0.01
D.O. %
0.1
Temp. °C
0.1
Accuracy
D.O.
±1.5% of full scale
Temp. °C
±0.5
Typical EMC
D.O.
±1.5% of full scale
Deviation
Temp. °C
±0.5
Calibration
D.O. single or double point at 0%
(HI 7040) and 100% (in air)
Temp. °C single point or double point
at 0.0°C and/or 50.0°C
Altitude Compensation
0 to 1,900 m (6,230')
Resolution
100 m (328')
Salinity Compensation
0 to 40 g/l
 Resolution
1 g/l
Temperature
Compensation
0.0 to 50.0°C (32 to 122°F)
Probe
HI 76407/2 with
2 meters (6.7') cable
  Logging Interval
1, 15, 30 seconds
or 1, 2, 5, 15, 30, 60, 120,
180 minutes
Computer
RS 232C (optoisolated)
Interface
Power
Power socket for 12 VCD
Environment
0 to 50°C (32 to 122°F);
95% RH
Dimensions
230x170x70 mm
 (9.1x6.7x2.7")
Weight
1 kg (2.2 lb.)
Response Time  The response time is approximately
20 seconds for a 95% reading at a constant temperature
of 25°C. The response time for low oxygen readings or
at low temperature is approximately 40 seconds. Allow
more time to obtain more accurate readings.
SPECIFICATIONS OF HI 964400