Hanna Instruments hi 93717 User Manual

Page of 2
Instruction Manual
HI 93717
Phosphate High Range
ISM
w w w . h a n n a i n s t . c o m
This Instrument is in
Compliance with the CE Directives
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
HI 93717 is warranted for two years 
against defects in
workmanship and materials when used for its intended purpose and
maintained according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the failure. If the repair is not covered
by the warranty, you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instruments, first obtain a Returned
Goods Authorization Number from the Customer Service department
and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with   directives.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note:
Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
The HI 93717 meter measures the phosphate (PO
4
3–
)
content in water and wastewater in the 0.0 to 30.0 mg/L
(ppm) range.
The meter uses an exclusive positive-locking system to ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in liquid and powder form and are
supplied in bottles and in packets. The amount of reagent
is precisely dosed to ensure maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONS
Range
0.0 to 30.0 mg/L
Resolution
0.1 mg/L
Accuracy
±1 mg/L ±4% of reading
Typical EMC
Deviation
±0.1 mg/L
Light Source
Light Emitting Diode @ 470 nm
Method
Adaptation of the 
Standard Methods
for the Examination of Water and
Wastewater (18
th
 edition), amino acid
method. The reaction between phosphate
and reagents causes a blue tint in the
sample
Light Detector
Silicon Photocell
Environment
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life
1 x 9 volt/40 hours
Auto-Shut off
After 10' of non-use
Dimensions
80 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight
290 g (10 oz.)
REQUIRED REAGENTS
Code
Description
Quantity
HI  93717A-0
Molybdate
10 drops
HI  93717B-0
Amino Acid Powder
1 packet
REAGENT SETS
HI 93717-01
Reagents for 100 tests
HI 93717-03
Reagents for 300 tests
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt
appears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
the maximum concentration readable (see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not inserted
correctly and an excess ambient light is
reaching the detector. If the cover is
properly installed, then contact your dealer
or the nearest Hanna Customer Service
Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery voltage
is getting low and the battery needs to
be replaced.
This indicates that the battery is dead
and must be replaced.
Note:
 once this indication is displayed,
the meter  will lockup. Change the
battery to restart.
V
Warranty
General Description
Preliminary Examination
Specifications
Display Code Guide