Hanna Instruments hi 98230 User Manual

Page of 21
9
8
PASSWORD PROTECTION
Setting of the GLP parameters (calibration alarm time-out, instru-
ment ID code, time and date) can be password protected. If password
is set to a value different from 0000 (factory setting), the user will be
asked to enter the password.
• Select the desired GLP parameter code.
• Enter the password by the arrow keys.
• Press the ALT and CFM keys to
confirm.
• If password is wrong the meter will return to the function selection
mode without any warning message.
• If password is correct, the meter provides access to the parameter.
PARAMETER SETTINGS
• Once the parameter code has been entered, the appropriate
message will scroll across the LCD for a few seconds.
• The current value of the selected parameter on the upper LCD
and the parameter code on the lower LCD will be displayed. The
first digit will blink if the parameter can assume continuous
values.  All the digits will blink if the parameter can assume only
a fixed set of values.
• Enter the new value using the arrow keys.
• Press ALT and CFM to confirm the value.
The following table lists the setup codes along with the description of
the specific setup items, their valid values and the factory settings
(default):
Code
Valid values
Default
00 Lot data printing/scrolling
00 16
00
01 Print lots data summary
02 Printer enable
On(enabled); Off(disabled) On
03 Logging interval
1,2,5,10,15,30,60,120,180
1
05 Log on demand delete
06 Timed data delete
Setup can be used to view data regarding instrument status (e.g.
battery charge) or GLP data (e.g. calibration date) or to view or print
the logged data. It also allows the user to change the meter
parameters (e.g. time) and to have access to stored data.
• To enter this mode ensure the meter is
not logging and then press the ALT and
FNC keys.
• The scrolling message "Insert the function code or press ALT - FNC
to escape" in the upper LCD and the indication "F 00" with the
first digit blinking in the lower LCD will be displayed.
• Enter the first digit of the code of the parameter you want to set
using the 
Ç or È key and pass to the next digit with Æ. The
second digit will start blinking.
• Enter the second digit using the 
Ç or
È key.
• Press ALT and CFM to confirm the code.
• If the entered code doesn't exist the "Err"
message will be displayed for a few
seconds and then the message "Insert
the function code or press ALT - FNC to
escape" will recommence scrolling in the
upper LCD.
SETUP MODE
SETUP MODE
SETUP MODE
SETUP MODE
SETUP MODE