Carbon Dioxide Electrode
 
 
3
 
EUTECH INSTRUMENTS PTE LTD. 
CARBON DIOXIDE GAS-SENSING ELECTRODE 
INSTRUCTION MANUAL 
 
GENERAL INSTRUCTIONS 
 
Introduction 
 
Eutech Carbon Dioxide Gas-Sensing Electrode is used to measure carbon dioxide, carbonate, and 
bicarbonate in aqueous solutions quickly, simply, accurately, and economically. With a flow-
through cap, the electrode can be used in flow-through applications. 
 
Required Equipment  
 
1. 
A pH/mV meter or an ion meter, either line operated or portable. 
 
2. 
Semi-logarithmic 4-cycle graph paper for preparing calibration curves when using the 
meter in the mV mode. 
 
3. 
A magnetic stirrer. 
 
4. 
Eutech Carbon Dioxide Gas-sensing Electrode, Code no. EC-C02-01. 
 
Required Solutions 
 
1. 
Deionized or distilled water for solution preparation. 
 
2. 
Eutech Carbon Dioxide Standard, 0.1M NaHCO
3
, Code no. EC-SCS-CO1-BT. To prepare 
this solution from your own laboratory stock, half fill a 1 liter volumetric flask with 
distilled water and add 8.40 grams of reagent-grade NaHCO
3
. Swirl the flask gently to 
dissolve the solid. Fill the flask to the mark with distilled water, cap, and upend several 
times to mix the solution. 
 
3. Eutech 
Carbon 
Dioxide 
Standard, 1,000 ppm NaHCO
3
 as CO2, Code no. EC-SCS-CO2-
BT.  To prepare this solution from your own laboratory stock, half fill a one liter 
volumetric flask with distilled water and add 1.91 grams of reagent-grade NaHCO
3
. Swirl 
the flask gently to dissolve the solid. Fill the flask to the mark with distilled water, cap, and 
upend several times to mix the solution. 
 
4. 
Eutech Carbon Dioxide Standard, 100 ppm NaHCO3 as CaCO
3
, Code no. EC-SCS-CO3-
BT. To prepare this solution from your own laboratory stock, half fill a one liter volumetric 
flask with distilled water and add 0.084 grams of reagent-grade NaHCO3. Swirl the flask 
gently to dissolve the solid. Fill the flask to the mark with distilled water, cap, and upend 
several times to mix the solution. 
 
5. 
Eutech Carbon Dioxide Buffer Solution, Code no. EC-ISA-CO1-BT. To prepare this 
solution from your own laboratory stock, half fill a 1 liter volumetric flask with distilled