eutech-instruments waterproof cyberscan con 400410 & tds 400 conductivitytds meter User Manual

Page of 56
Instruction Manual 
 
CON 400/410 
- 16 - 
5.2 Manual 
Temperature 
Compensation 
IMPORTANT:  For manual compensation, you must deactivate the automatic temperature 
compensation feature.  
 
5.2.1 
Selecting Manual Temperature 
Compensation 
To select between Automatic Temperature Compensation 
(ATC) and Manual Temperature Compensation in the 
SETUP program P4.3. Meter default is ATC on.  
 
From measurement mode 
1) Press 
SETUP key to enter Set Up mode.  
2) Press 
the 
MI/
S or MR/T keys to scroll through 
subgroups until you view parameter P4.0.  
3) Press 
the 
ENTER key three times to select parameter 
4.3. The upper display shows “ATC” and the lower 
display shows “P4.3”.  
4) Press 
the 
ENTER key again. The upper display shows 
“ATC” and the lower display shows “YES” or “NO”.  
5) Press 
the 
MI/
S or MR/T to select between Automatic 
or Manual Temperature Compensation feature.  
•  YES = Automatic Temperature Compensation 
activated. 
•  NO = Manual Temperature Compensation 
activated. 
6) Press 
the 
ENTER key to confirm selection and to 
return to the subgroup menu. Press the CAL/MEAS 
key to return to measurement mode.   
 
NOTE:  
Manual Temperature Compensation should only be 
selected if the ATC fails. For non-compensated 
measurements, change the temperature coefficient to 
0.0% (see section 7.6.1) 
 
P4.0 Configuration Setup 
Selecting Automatic or Manual 
Temperature Compensation