eutech-instruments waterproof cyberscan con 400410 & tds 400 conductivitytds meter User Manual

Page of 56
Instruction Manual 
 
CON 400/410 
- 28 - 
7.3 
P2.0: Viewing calibration data  
This mode lets you recall previous calibration data, which 
helps you know when to re-calibrate your meter.  This is 
a “view only” mode. See Figure 24. 
 
From measurement mode: 
1. Press 
the 
SETUP key to enter Set up mode.  
2. Press 
the 
MI/
S or MR/T  keys to scroll through 
subgroups until you view parameter P2.0. 
3. Press 
ENTER key repeatedly to view previous 
calibration data. The meter will first display the 
calibration point, and then display the date and time 
of calibration to help you comply with the Good 
Laboratory Practice (GLP) standards.  
4. 
When you have scrolled through all calibration data, 
you will automatically return to the subgroup menu. 
Press CAL/MEAS key to return to measurement 
mode.  
 
 
 
 
 
 
NOTES: 
If there is no previous calibration data at a particular point, 
the primary display will show “ ----“.  
 
CON 410 meter only:  If you entered Set Up mode from 
Conductivity measurement mode, calibration data will be 
in 
µS or mS. If you entered Set Up mode from TDS 
measurement mode, calibration data will be in ppm or ppt.  
 
 
Figure 24 - View calibration data 
and its relevant time and date