eutech-instruments waterproof cyberscan con 400410 & tds 400 conductivitytds meter User Manual

Page of 56
Instruction Manual 
 
CON 400/410 
- 42 - 
1 0   E R R O R  M E S S A G E S  
LCD Display 
Indicates 
Cause 
Solution 
Err annunciator  
Unrecognised input 
from keypad 
Wrong input in selected 
mode.  
Release key. Select valid 
operations depending on 
mode.  
CAL & Err 
annunciators on / 
Buffer and 
electrode 
indicators blink.  
Calibration error.  
Wrong value input at 
calibration.  
Dirty probe.  
Check your input value, 
clean probe.  
See Calibration sections or 
Probe Maintenance 
section.  
Battery indicator 
blinks 
Low battery level.  
Need new batteries or 
battery connection is 
bad.  
Clean battery contacts. 
Replace batteries with fresh 
ones, noting polarity.  
Err 1 (in primary 
display) 
Memory write error.  
Hardware failure.  
Turn meter on and off 
again.  If message persists, 
return unit*.  
Err. 2 (in primary 
display) 
Memory checksum 
error.  
Hardware failure.  
Turn meter on and off 
again.  If message persists, 
return unit*. 
Err. 3 (in primary 
display) 
A/D converter error.  
Hardware error.  
Turn meter on and off 
again.  If message persists, 
return unit*. 
Err. 4 (in primary 
display) 
Keypad error.  
One or more keys on 
the keypad are stuck or 
fault in keypad.  
Turn meter on and off 
again.  If message persists, 
return unit*. 
 
* See “Warranty” and “Return of Items” on pages 49 and 50.  
If an error message appears in the primary display (the 
upper row of larger digits), switching off the meter and 
switching it on again may eliminate the error message. 
Refer to diagram on right.  
 
If error persists, or the meter shows incorrect values, return the meter.  
For a complete diagram of the display see page 2.