9.9  Controller Reset – P6.0 
This parameter lets you reset the controller. All pH and ORP calibration data will be 
cleared, and all Setup parameters revert to factory default values. Temperature 
mode (ATC or MTC) and calibration data remains unchanged. 
P6.0
RsT
n0
yeS
P6.1
 
Press ▲ and ▼ keys together (ESCAPE) at anytime, to leave the Setup mode 
and return to Measurement mode. 
 
P6.0: Controller reset 
1. 
Key in the password “022” as per procedure in Section 5. 
2. 
Press ▲ key five times. The screen will scroll P6.0 and RST. Press ENTER. 
3. 
The screen will scroll P6.1, then NO. 
4. 
Press  ▲ or ▼ keys to toggle between YES and NO. Pressing YES will 
clear all calibration data and setup parameters! 
5. 
Press ENTER to confirm selection and return to general Setup mode.  
Press ▲ and ▼ keys together (ESCAPE) to return to measurement mode.