eutech-instruments cadmium epoxy User Manual

Page of 20
Cadmium Electrode 
 
Instruction Manual
 
 
18
TROUBLESHOOTING HINTS 
 
Symptom           
       
Possible Causes          
 
 Next Step 
Out of Range          
 
defective meter       
 
 
check meter with 
Reading                                    
 
 
   
 
shorting strap 
                                                
 
 
 
 
(see meter instruction manual) 
 
 
 
 
      
defective electrode      
 
check electrode operation 
                       
                         
 
electrodes not      
  
 
unplug electrodes 
                           
 
plugged in properly      
 
and reset 
 
 
 
              
reference electrode 
 
 
be sure reference 
                        
 
not filled 
 
 
            electrode is filled 
 
 
 
                   
air bubble on    
 
 
remove bubble by 
                    
 
 
membrane          
 
 
redipping electrode 
 
 
 
                  
electrodes not   
  
 
put electrodes 
                         
 
in solution  
 
 
            in solution 
 
Noisy or Unstable        
defective meter       
 
 
check meter with 
Readings (readings                             
 
 
 
shorting strap 
continuously or 
rapidly changing)  
 
air bubble on       
 
 
remove bubble by  
                       
 
membrane              
 
 
re-dipping electrode 
 
  
 
                
electrode exposed       
 
soak electrode in 
                       
 
to interferences      
 
 
cadmium standard 
 
                                      
defective electrode  
 
 
replace electrode 
 
 
 
               
ISA not used       
 
 
use recommended ISA 
 
 
 
             
meter or stirrer    
 
 
ground meter or 
                       
 
not grounded     
 
 
stirrer 
 
 
Drift (reading        
 
samples and standards   
 
allow solutions to come 
slowly changing          
at different         
 
 
to room temperature 
in one direction)   
 
temperatures              
 
before  measurement 
 
 
 
                    
electrode exposed        
 
check section  
                    
 
 
to complexing agents   
 
entitled 
Precipitation 
 
 
            
                                                        and Complexation