8 .   T R O U B L E S H O O T I N G  
Problem Cause 
Solution 
No display when 
turned on 
1. 
Batteries not in place. 
 
2. 
Batteries not in correct polarity 
(+ and -). 
3. Weak 
batteries 
1. 
Check that batteries are in place 
and making good contact.  
2. 
Reinsert batteries with correct 
polarity.  
3. 
Attach optional AC adapter (only 
for Standard PC 10 meter) or 
replace batteries.  
Unstable readings 
1. 
Air bubbles in probe.  
2. 
Dirty probe.  
3. 
Probe not deep enough in 
sample.  
4. 
External noise pickup or 
induction caused by nearby 
electric motor.  
5. 
Broken probe.  
6. Temperature 
is 
changing 
 
1. 
Tap probe to remove bubbles.  
2. 
Clean the probe and recalibrate.  
3. 
Make sure sample entirely covers 
the probe sensors.  
4. 
Move or switch off interfering 
motor. 
 
5. 
Replace probe.  See page 21.  
6. 
Wait for temperature reading to 
stabilise. 
“OR” on upper 
display 
1. 
Probe is shorted.  
 
2. 
Probe is in an out-of-range 
solution.  
3. 
Broken probe.  
1. 
Test probe.  Make sure probe is 
fully connected to meter.  
2. 
Use different solution.  
 
3. 
Replace probe. See page 21.  
Temperature 
reading erratic or 
lower display reads 
“OR”.  
1. 
Temperature sensor is dirty.  
 
2. 
Temperature of solution is out 
of range.  
1. 
Clean temperature sensor with 
isopropyl alcohol.  
2. 
Heat or cool solution.  
Slow response 
1. 
Dirty / Oily probe.  
1. 
Clean probe. See “Probe Care & 
Maintenance”, page 19.