eutech-instruments chloride epoxy User Manual

Page of 21
Instruction Manual 
 
Chloride Electrode 
 
19
TROUBLESHOOTING HINTS 
 
Symptom                 Possible Causes          
 
 
Next Step 
Out of Range            defective meter           
 
 
check meter with shorting 
Reading                                       
 
 
 
 
strap (see meter  
                                                
 
 
 
 
instruction manual) 
 
                         
defective electrode 
 
 
       
check electrode operation 
                       
                         
electrodes not          
 
 
 
plug electrode into position  
   plugged 
in 
properly 
   and 
reset 
meter 
 
  
 
 
                         
reference electrode 
 
 
 
clean junction 
                         
junction clogged 
 
   static 
electricity    wipe 
plastic 
meter 
face 
with 
 
 
 
 
 
 
 
 
 
detergent solution 
 
   meter 
or 
stirrer     ground 
meter 
or 
stirrer
 
 
   not 
grounded 
 
 
Noisy or Unstable      defective meter           
 
 
check meter with 
Readings (readings                              
 
 
 
shorting strap 
continuously or 
rapidly changing)       air bubble on membrane           
 
remove bubble by  
                        
 
 
             
 
 
re-dipping electrode 
 
                         
ISA not used              
 
 
use recommended ISA 
 
                         
meter or stirrer        
 
 
 
ground meter or 
                         
not grounded              
 
 
stirrer 
 
                         
defective electrode       
 
 
replace electrode 
 
                         
electrode exposed        
 
 
soak electrode in 
                         
to interferences          
 
 
chloride standard 
 
Drift (reading         
samples and standards   
 
 
allow solutions to come to  
slowly changing         at different temperatures         
 
room temperature before 
in one direction)                                                   
 
 
measurement 
 
                                    incorrect reference       
 
 
use recommended 
                         
filling solution        
 
 
 
filling solution in outer 
 
 
 
 
 
 
 
 
 
of reference electrode 
  
 
   membrane 
dirty 
or 
etched 
  polish 
membrane