calibration failure for 0% Saturation slope calibration; the meter may have 
exceeded the limit of 10% of the factory calibrated absolute value. The sensor 
may have to be re- conditioned. Also, note that the DO 6 meter will not perform 
0% or 100% Saturation calibration for absolute value ranging from 10.1% to 
49.9%. The meter also will not perform the mg/L (ppm) Concentration 
calibration for an absolute value of less than 2.00 mg/L (ppm) or calibration 
which is out of its window tolerance of +/- 40% from the factory default value. 
An error message of “Err.1” will be indicated and will return the meter to the 
measurement mode.  
All new calibration values will automatically override the existing data. It is 
recommended to calibrate the meter periodically and or if it is suspected to be 
inaccurate. 
Always rinse the probe with either DI water or rinse solution before and after 
each calibration/sample measurement. When calibrating in air, make sure that 
any water droplets from the probe’s membrane are removed. 
For details please refer to Section 8 - PROBE CARE AND MAINTENANCE
Note: The DO6 factory calibrated default value is in respect to 760mm Hg or 
101.3 kPa barometric pressure (sea level).
 
To set the barometric pressure to 
different value, see Section 5.4.1 Pressure Setting Adjustment.